fully automatic  cleaning and disinfection equipment CCL260916

Product info

Packaging Specifications

Core Parameters

Description

Summary of CCL260916 Fully‑Automatic Cleaning and Disinfection Machine

Product Overview

Vertical automatic medical‑device cleaning‑disinfection equipment for reusable medical instruments. It delivers standardized automated cleaning to prevent cross‑infection, available in 260 and 360 models.

Main Features

  1. Cleaning & drying: Imported high‑flow circulation pump; patented omni‑directional rotating spray arm without cleaning dead zones; patented leak‑proof interface for cleaning chamber and racks. 120 ℃ hot‑air drying with H13 HEPA filter (filtration precision ≤0.3 μm). Zero‑waiting full‑automatic workflow: enzyme washing, double rinsing, disinfection/oiling, drying.
  2. Control & data record: Real‑time temperature and A0‑value monitoring on touch screen. ≥6 factory‑preset programs plus ≥26 customizable programs. Automatic precise reagent injection (13.5 mL/s, adjustable volume). Built‑in needle printer records runtime, program name, operator ID, phase temperature parameters, A0 value and total running counts.
  3. Safety configuration: Safety self‑check and intelligent fault diagnosis; infrared anti‑pinch‑hand protection; double‑door interlock; imported core components including circulation pump, micro‑switch and photoelectric switch.
  4. Application scope: Surgical cutters, hemostatic forceps, bowls, trays, drums; compatible with respiratory tubes, anesthesia tubes, minimally‑invasive instruments and plastic accessories with matched racks and programs.

Model Difference

  • Model 260: 7‑inch touch screen, manual drop‑down door; power 20 kW, weight 325 kg; power supply AC 380V 50Hz ≥40A; water consumption 25 L per step; max cycle time ≤50 min.
  • Model 360: 10.1‑inch touch screen, automatic lifting door; higher power and weight; power supply AC 380V 50Hz ≥60A; lower maximum cycle time down to ≤38 min, higher water consumption per step.

Operating environment: Temperature 5‑40 ℃, relative humidity ≤75%, atmospheric pressure 70‑106 kPa; minimum maintenance clearance: 500 mm.