Summary of CCL‑AP240 Fully‑Automatic Specific Protein Analyzer (Closed‑System)

Manufacturer: China Care Medical Equipment Co., Ltd.


Product Overview

CCL‑AP240 is a closed‑system fully‑auto specific‑protein analyzer. It adopts nephelometry, colorimetry and turbidimetry. It delivers throughput up to 100 tests per hour, supports diverse sample matrices including whole blood, serum, plasma, urine and CSF, covering renal‑function, rheumatism, inflammation, immunoglobulin, multiple myeloma‑related markers and other clinical panels.


Main Features


  1. High‑efficiency performance: 100 tests/hour throughput; continuous‑loading capability; independent translational STAT emergency‑sample position. Auto‑antigen‑excess detection, automatic dilution & retest (dilution ratio from 1:1 up to 1:40 000).
  2. Reliable detection quality: Advanced nephelometry improves sensitivity & precision. Three‑step cuvette cleaning reduces cross‑contamination. Equipped with liquid‑level detection, probe clogging & dirty‑cuvette monitoring. On‑board 24‑h reagent cooling preserves reagent stability.
  3. Cost‑optimized design: 32 reusable cuvettes plus one independent dilution station. Reagent kits include calibrators and QC materials. One reagent kit works for both blood and urine specimens; long‑life light‑source; low water consumption (1.2 L/h).
  4. User‑friendly workflow: Built‑in barcode scanning for samples and reagents; one‑touch QC & multi‑point calibration via magnetic‑card calibrator. 10.4‑inch touch‑screen; bi‑directional LIS / HIS communication; local storage for up to 150 000 sample results.


Main Technical Specifications



ItemSpecification
Detection PrincipleNephelometry, colorimetry, turbidimetry
ThroughputUp to 100 T/H
Positions40 reagent/sample positions; 32 reusable cuvettes; 1 independent dilution position
Sample TypesWhole blood, serum, plasma, urine, CSF
Dilution Range1:1‑1:40 000
Incubation Temperature37 ± 0.5 ℃
WavelengthNephelometry:630‑690 nm; Colorimetry/Turbidimetry:340‑700 nm multiple wavelengths
CalibrationMagnetic‑card & calibrator, multi‑point calibration
Storage150 000 sample results
Interfaces6×USB, 1×RJ45, 1×serial port
Water Consumption1.2 L/h
Dimension650 × 530 × 650 mm
Net Weight64 kg