The REM system continually monitors patient impedance levels and deactivates the generator if a fault in the patient/return electrode contact is detected, which minimizes the danger of burning incidents.
This proprietary technology delivers optimal clinical effects through continuous synchronization of current and voltage. It samples the current and voltage 450,000 times per second, enabling it to respond to tissue impedance changes in less than 10 milliseconds. This ensures the machine achieves optimum energy output levels quicker and more accurately, delivering only the precise voltage needed safely to each tissue type.
The machine enables permanent sealing of blood vessels with a diameter up to 7mm with this technology under bipolar coagulation.
Once switched on, the system performs a comprehensive internal test.
表格
| Working Mode | Sub-mode | Max. Output Power | Rated Load |
|---|---|---|---|
| Monopolar Cut | Pure Cut | 150 W | 500 Ω |
| Blend Cut I | 120 W | 500 Ω | |
| Blend Cut II | 100 W | 500 Ω | |
| Monopolar Coagulation | Fulgurate Coagulation | 120 W | 500 Ω |
| Spray Coagulation | 100 W | 500 Ω | |
| Bipolar Coagulation | Standard Coagulation | 80 W | 100 Ω |
| Ligasure Vessel Sealing | - | 150 W | 25 Ω |