- Description
Specifications:
Reagent
Net weight: 8KG
Gross weight:12KG
- Technology:ISE(Ion Selectivity Electrode),Measurement of Tco2 by pressure method.
- Sample:Serum/Plasma/Whole Blood/Urine/CSF
- Sample Volume:80-150ul
- Reagent Volume:1ml
- Throughput:60 samples/hour
- Printer:Inner Thermal Printer
- Display:Large LCD
- Storage:Up to 1000 tests
- Interface:RS-232
- Working Conditions:Temperature:15-30℃
- Humidity:20%-80%
- Power Requirement:220AV±10%,50/60 Hz,80W Electrode
- ISE electrode used for Ge300 series Electrolyte Analyzer
- Parameters:K+,Na+,Cl-,Ca++,Li+,PH,Ref,Tco2
Item | Linearity(mmol/L) | Slope | Lifespan |
K+ | 0.01-99.00 | 58.0 | 12 months |
Na+ | 1.0-999.0 | 58.0 | 8 months |
Cl- | 1.0-999.0 | 56.0 | 8 months |
Ca++ | 0.01-9.9 | 28.0 | 8 months |
Li+ | 0.01-9.9 | 56.0 | 8 months |
PH | 3.50-10.08 | 58.0 | 12 months |
Ref | / | / | 36 months |
Reagent
- Standard Reagent A&B 390/90 ml
- Standard Reagent A&B&C 390 ml
- QC Solution 80ml
- Cleaning solution 80 ml
- Electrode filler 10 ml
- Ref.electrode refiller 10 ml
Model | Test items |
CCL-300A | K+,Na+ |
CCL-300B | K+,Na+,Cl- |
CCL-300C | K+,Na+,Cl-,iCa2+,TCa,pH |
CCL-300D | K+,Na+,Cl-,iCa2+ |
CCL-300E | K+,Na+,Cl-,Li+ |
CCL-300F | K+,Na+,Cl-,iCa2+,pH,TCO2,TCa,AG,KT |