Electrosurgical Unit


Technical Features


REM (Return Electrode Monitoring)

The REM system continually monitors patient impedance levels and deactivates the generator if a fault in the patient/return electrode contact is detected, which minimizes the danger of burning incidents.


Real-time Monitoring and Instance Response System for Tissue Density

This proprietary technology delivers optimal clinical effects through continuous synchronization of current and voltage. It samples the current and voltage 450,000 times per second, enabling it to respond to tissue impedance changes in less than 10 milliseconds. This ensures the machine achieves optimum energy output levels quicker and more accurately, delivering only the precise voltage needed safely to each tissue type.


Ligasure Vessel Sealing

The machine enables permanent sealing of blood vessels with a diameter up to 7mm with this technology under bipolar coagulation.


Automatic Self-Test

Once switched on, the system performs a comprehensive internal test.


Technical Specification


Working Modes

表格


Working ModeSub-modeMax. Output PowerRated Load
Monopolar CutPure Cut150 W500 Ω
Blend Cut I120 W500 Ω
Blend Cut II100 W500 Ω
Monopolar CoagulationFulgurate Coagulation120 W500 Ω
Spray Coagulation100 W500 Ω
Bipolar CoagulationStandard Coagulation80 W100 Ω
Ligasure Vessel Sealing-150 W25 Ω