The intuitive design and user-friendly set-up of the Electrosurgical Generator (diathermy) for use in the modern operating room, it features both Monopolar and bipolar functions to satisfy all the surgeries demands with safety, flexibility, reliability and convenience.
Return electrode contact quality monitoring system (REM). The REM system continually monitors patient impedance levels and deactivates the generator if a fault in the patient/return electrode contact is detected, which minimize the danger of burning incidents.
Language options: English, Spanish, Portuguese, Turkish and etc.
General Surgery; Gastroenterology; Dermatology; Vascular Surgery; Obstetrics and Gynecology; Heart/Thoracic Surgery; ORL/ENT; Minimally Invasive Surgery (MSI); Cerebral surgery; Neurosurgery; Orthopedics & Plastic Surgery; Trans Urethral Resection (TUR) and etc.
| Working Mode | Sub-mode | Maximum P-P Voltage | Rated Load | Max. output power | Crest Factor (Typical) |
|---|---|---|---|---|---|
| Monopolar Cut | Pure | 1900 | 500Ω | 400W | 1.6 |
| Monopolar Cut | Blend 1 | 2300 | 500Ω | 250W | 2.5 |
| Monopolar Cut | Blend 2 | 2500 | 500Ω | 200W | 2.8 |
| Monopolar Cut | Blend 3 | 2800 | 500Ω | 150W | 2.9 |
| Monopolar Coag | Fulgurate coag | 2800 | 500Ω | 120W | 2.9 |
| Monopolar Coag | Spray coag | 7000 | 500Ω | 100W | 6.2 |
| Bipolar Coag | Standard Coag | 550 | 100Ω | 70W | 1.6 |