The intuitive design and user-friendly set-up of the Electrosurgical Generator (diathermy) for use in the modern operating room, it features both Monopolar and bipolar functions to satisfy all the surgeries demands with safety, flexibility, reliability and convenience.
Return electrode contact quality monitoring system (REM). The REM system continually monitors patient impedance levels and deactivates the generator if a fault in the patient/return electrode contact is detected, which minimize the danger of burns.
Language options: English, Spanish, Portuguese, Turkish and etc.
General Surgery; Gastroenterology; Dermatology; Vascular Surgery; Obstetrics and Gynecology; Heart/ Thoracic Surgery; ORL/ENT; Minimally Invasive Surgery (MSI); Cerebral surgery; Neurosurgery; Orthopedics & Plastic Surgery; Trans Urethral Resection (TUR) and etc.
Under water cut socket (Endo cut)
HV argon coagulation system
Optimum Smoke Evacuation System-Clean air in the operation theatre
Main Frequency: 512 KHz
Power Supply: 110V-220V; 50-60Hz
Weight: 8.00KG for Stainless Steel model; 6.00KG for Injection model
Dimension: 413214cm
Monopolar: < 150mA
Bipolar: < 60mA
Normal Polarity, Intact Chassis Ground: <10µA
Normal Polarity, Ground open: <50µA
Reverse Polarity, Ground Open: <50µA
Sink Current, 140V Applied, All inputs: <50µA
| Working Mode | Sub-mode | Maximum P-P Voltage | Rated Load | Max. output power | Crest Factor (Typical) |
|---|---|---|---|---|---|
| Monopolar Cut | Pure | 1900 | 500Ω | 300 | 1.6 |
| Monopolar Cut | Blend 1 | 2300 | 500Ω | 250 | 2.5 |
| Monopolar Cut | Blend 2 | 2500 | 500Ω | 200 | 2.8 |
| Monopolar Cut | Blend 3 | 2800 | 500Ω | 150 | 2.9 |
| Monopolar Coag | Fulgurate | 2800 | 500Ω | 120 | 2.9 |
| Monopolar Coag | Spray | 7000 | 500Ω | 100 | 6.2 |
| Bipolar Cut | Standard Cut | 700 | 100Ω | 70 | 1.6 |
| Bipolar Coag | Standard Coag | 550 | 100Ω | 70 | 1.6 |
no comments