This integrated neonate bilirubin phototherapy equipment combines an incubator with dual phototherapy systems (upper and nether cases), providing a controlled environment for effective hyperbilirubinemia treatment. It features microcomputer control, dual light source options, and a range of safety and monitoring functions.
| Item | Details |
|---|---|
| Power Requirement | AC220V–230V/50Hz or AC110–120V/50–60Hz, 600VA |
| Control Mode | Air mode and baby mode controlled by micro-computer |
| Temperature Control Range | 25°C–34°C |
| Skin Temp Sensor Display Range | 5°C–65°C |
| Displayed vs Actual Value Correlation | ≤0.8°C (stable temperature) |
| Skin Temp Sensor Accuracy | ≤0.2°C |
| Bed Temp Uniformity | ≤2.0°C |
| Noise Level (Above Bed) | ≤55dB(A) |
| Failure Alarms | Over-temperature, Deviation, Sensor failure, Fan failure, Check alarm |
| Total Irradiance (Above Case) | ≥1.5mW/cm² (blue fluorescent lamp); ≥2.0mW/cm² (Blue LED) |
| Total Irradiance (Nether Case) | ≥3.0mW/cm² (Blue LED) |
| Average Bilirubin Radiation (Above Case) | ≥1.4mW/cm² (blue fluorescent lamp); ≥1.5mW/cm² (Blue LED) |
| Average Bilirubin Radiation (Nether Case) | ≥2.5mW/cm² (Blue LED) |
| Irradiance Uniformity | >0.4 |
| Dominant Blue Light Wavelength | 400nm–550nm |