Product info

Packaging Specifications

Description

Neonate Bilirubin Phototherapy Equipment

This integrated neonate bilirubin phototherapy equipment combines an incubator with dual phototherapy systems (upper and nether cases), providing a controlled environment for effective hyperbilirubinemia treatment. It features microcomputer control, dual light source options, and a range of safety and monitoring functions.


Brief Introduction


Standard Configuration


Optional Configuration

ItemDetails
Power RequirementAC220V–230V/50Hz or AC110–120V/50–60Hz, 600VA
Control ModeAir mode and baby mode controlled by micro-computer
Temperature Control Range25°C–34°C
Skin Temp Sensor Display Range5°C–65°C
Displayed vs Actual Value Correlation≤0.8°C (stable temperature)
Skin Temp Sensor Accuracy≤0.2°C
Bed Temp Uniformity≤2.0°C
Noise Level (Above Bed)≤55dB(A)
Failure AlarmsOver-temperature, Deviation, Sensor failure, Fan failure, Check alarm
Total Irradiance (Above Case)≥1.5mW/cm² (blue fluorescent lamp); ≥2.0mW/cm² (Blue LED)
Total Irradiance (Nether Case)≥3.0mW/cm² (Blue LED)
Average Bilirubin Radiation (Above Case)≥1.4mW/cm² (blue fluorescent lamp); ≥1.5mW/cm² (Blue LED)
Average Bilirubin Radiation (Nether Case)≥2.5mW/cm² (Blue LED)
Irradiance Uniformity>0.4
Dominant Blue Light Wavelength400nm–550nm

Environmental Requirement


Product Packaging

User review