Reagent System for hematology cell counter,
Can be customized into open system, which can be suitable for international brand, Like Mindray, Backmen and others.
(hematology cell counter Control System)
Test principle: electrical resistance for counting and SFT method for HGB
Test flow: fully auto and the blood don’t need dilute or other process
Clot clear: high voltage burn and high pressure flush
Parameter: 3 part differentiation of WBC,20 parameters + 3 histograms
Sample volume: 9.8μl peripheral blood or venous blood
Test channel: two channels
Throughput: 60 samples/hour
Language: English,French, Russsia and others
(hematology cell counter Control System Calibration)
Calibration mode: auto mode and manual mode are selectable, provide original calibrator and control
Reference range type: newborn,children,male,female and universal
Reagent: provide original reagent
Memory: save more than 30000 sample’s detail results including histograms
Display: Large colour LCD, show all parameters and histograms at same screen
Input and output: keyboard,mouse,Parallel Port,RS232 serial port, can be connected to computer
Printer: Inner thermal printer,can be connected to external computer and printer
Working Environment:
Temperature:18℃~35℃; Humidity:40%RH-80%RH
Voltage: AC 175 V~265V,Auto adapt according to the input voltage
Dimension: 300 mm×400 mm×440mm(L×W×H)
Net weight: 18Kg
(Auto Hematology Analyzer Menu)
| Item | Precision, CV |
|---|---|
| WBC | ≤2.0 % |
| RBC | ≤1.9 % |
| HGB | ≤1.9 % |
| MCV | ≤0.4 % |
| PLT | ≤4.0 % |
| Carry over | WBC≤2 %,RBC≤1 % |
Sample volume: normal mode: 9.8μl, dilution mode: 20μl
Aperture diameter: WBC: 100μ m, RBC,PLT: 70μ m
Throughput: 60 samples/hour
Clot clear: high voltage burn and high pressure flush
Calibration mode: auto mode and manual mode are selectable, provide original calibrator and control
Display: Large colour LCD, show all parameters and histograms at same screen
Memory: save more than 30000 sample’s detail results including histograms
Input and output: keyboard,mouse,Parallel Port,RS232 serial port, can be connected to computer
Printer: Inner thermal printer,can be connected to external computer and printer
WBC Lymph# Mid# Gran#
Lymph% Mid% Gran% RBC
HGB HCT
MCV MCH MCHC RDW-SD RDW-CV PLT MPV PDW PCT P-LCR
WBC Histogram RBC Histogram PLT Histogram
If this hematology cell counter model can't satisfy your request, please click following links to check more other models.