This is a high-performance low-pressure ion-exchange liquid chromatography system, the gold standard for HbA1C testing, designed for accurate, reproducible results.
| Parameter | Specification |
|---|---|
| Testing method | Low-pressure ion-exchange liquid chromatography |
| Test item | Glycated hemoglobin (HbA1C) |
| Measuring range | 3% – 18% |
| Precision | CV ≤ 3% |
| Accuracy | ≤ 1.50% |
| Testing time | 4 minutes (reporting); full process 6 minutes (including cleaning and recovery) |
| Sample type | Venous blood (EDTA anticoagulated) or peripheral finger blood |
| Sample volume | 5 µL whole blood |
| Photometer | 415 nm LED integral flow colorimeter |
| Sampling mode | 20-position auto sample turntable |
| Testing mode | Batch auto-test; single emergency test |
| Calibration | 2-point or 4-point, manual or auto; with high/low calibrators |
| Thermostatic control | 25 ± 0.5 °C (column and reagents) |
| Reagent life | 1 column/reagent set = 300 tests |
| Output display | 320×240 LC graphic display (real-time curve) |
| Printer | 58 nm thermal printer (prints curve and report) |
| Report formats | IFCC concentration, NGSP area percentage, ADAG average glucose |
| Data storage | 1000 test reports (including curves) |
| Interface | RS232 (connects to HIS/LIS) |
| Power supply | AC 220 V ±22 V, 50 Hz, 65 VA |
| Dimensions | 408 × 390 × 230 mm |
| Weight | 8 kg |
| Working environment | Temperature 12–30 °C, relative humidity ≤ 85% |
no comments